Remote Interface Reference
IEEE-488 Configuration
the parameters under the Net Setup submenu (see sections 3.10.5.2,
3.10.5.3 and 3.10.5.4). To make the values dynamically assigned by a
network DHCP server, set IP Address Assignment to DHCP (see section
The system name (also known as
host name
or
computer name
), can be set
using remote communications (either Ethernet or RS-232); it cannot be
edited using the front panel keypad.
All network parameters (even those assigned by a DHCP server) can be
viewed using the Net Settings submenu (see section 3.10.4).
4.4.1
Ethernet Connector
The Model 430 Programmer uses a standard RJ-45 jack for Ethernet
communications. The Ethernet jack pinout is fully documented in Table A-
8 on page 160 in the
Appendix
.
4.4.2
Termination Characters
All commands and queries are transmitted and received as ASCII values
and are case insensitive. The Model 430 Programmer always transmits
<CR><LF>
(a
carriage return
followed by a
linefeed
) at the end of an
Ethernet transmission. The Model 430 can accept
<CR>
,
<LF>
,
<CR><LF>
, or
<LF><CR>,
or a semicolon (
;
) as termination characters
from an external computer.
Summary of Contents for 05100PS-430-601
Page 2: ......
Page 10: ...x Rev 5 List of Figures ...
Page 12: ...xii Rev 5 List of Tables ...
Page 18: ...xviii Rev 5 Foreword Safety Summary ...
Page 30: ...12 Rev 5 Introduction Operating Characteristics ...
Page 42: ...24 Rev 3 Installation Power Up Procedure ...
Page 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Page 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Page 168: ...150 Rev 5 Service Return Authorization ...
Page 226: ...208 Rev 5 Index ...