Connecting the system

Preparation for use

Turn the page

Ambu

®

 aScope

 3 Family

- Quick Guide for intubation and single lung ventilation

The aScope

TM

 3 Slim, aScope

TM 

3 Regular and aScope

TM 

3  

Large are the perfect choice for flexible fibreoptic intubation 
as they are ready to use whenever you need them. The 
single–use and ready–to–go concept simplifies flexible 
fibreoptic intubation and improves patient safety.
aScope

TM

 3 Slim, aScope

TM 

3 Regular, and aScope

TM 

3 Large are 

sterile and single-use flexible scopes with a 1.2 mm, 2.2 mm 
or 2.8 mm working channel respectively.
For intubation the  aScope

TM

 3 Slim can be used with ET tubes 

size 5 and up, aScope

TM 

3 Regular with ET tubes size 6 and up 

and aScope

TM

 3 Large with ET tubes size 7 and up. 

For single lung ventilation aScope 3 Slim can be passed 
through DLT’s of FR 37 or greater or BB of sizes 5–9.

   Ensure that the selected ET tube  

or DLT/BB is compatible with the 
aScope

TM

 3 chosen.

  Lubricate the insertion cord with  

a medical grade lubricant to  
ensure the lowest possible friction. 
Do not get lubrication or any other 
contaminant on the tip of the 
aScope

TM

 3.

  Attach the suction tube to the suction 

connector and ensure that the 
suction works when pressing the 
suction button

  Fluids can be instilled through the 

working channel by inserting a 
syringe into the working channel  
port at the top of the aScope

TM

 3.  

When using a Luer Lock syringe,  
use the included introducer. 

  If you are using a preformed ET  

tube unbend it before inserting  
the cord of the aScope

TM 

3 through it.

  Antifogging agent is not required  

with this device.

  Confirm that the bending section 

functions smoothly and correctly by 
carefully sliding the bending control 
lever forward and backward in each 
direction.

  Confirm that the bending section 

returns to a straight position.

  Connect the aScope

TM

 3 to the 

aView

TM

 and verify that a live video 

image appears on the screen.

  Always have an additional aScope

TM

 3 

at hand as a precaution.

1.

Functions

Warning: Read Instructions For Use before using the aScope

TM

 3 and aView

TM

1. & 2. Connect aScope

TM

 3 to the aView

TM

  by 

plugging the aScope

TM

 3 connector – white with 

blue arrow - into the corresponding blue female 
connector on the side of aView

TM 

.

3. Push the ON/OFF
button on the aView

TM

 

to start up the system.

4. The real time image
will appear on aView

TM

after only a few seconds
and the aScope

TM

 3 

system is ready to use