![AMALGAMATED INSTRUMENT RM4-IV Operation And Instruction Manual Download Page 31](http://html1.mh-extra.com/html/amalgamated-instrument/rm4-iv/rm4-iv_operation-and-instruction-manual_2915944031.webp)
dISP
- display mode. The retransmission output will follow whatever value is on the 7 segment
display. For example if the remote input is set to
tARE
then the 7 segment and retransmission
output will indicate the tared value and both will also be changed if the remote input toggles the
displays between
nEtt
and
GROS
. If the
REC
or
dGOP
function had been set to
tARE
then
the retransmission output would not respond to the
GROS
toggle.
5.36
Second analog output mode
Display:
REC2
Range:
LiuE, tARE, P.HLd, d.HLd, Hi, Lo
or
dISP
Default Value:
LiuE
Analog output mode for second analog output. Operates in the same manner as the first analog
output, see function
5.37
Low overrange visual warning limit value
Display:
Lo dISP
Range:
Any display value or
OFF
Default Value:
OFF
Low overrange limit value - the display can be set to show an overrange message if the display
value falls below the
Lo dISP
setting. For example if
Lo dISP
is set to
50
then once the
display reading falls below
50
the message
-or-
will flash on and off or the display value will
flash on and off instead of the normal display units (see
dISP
function
). This message can
be used to alert operators to the presence of an input which is below the low limit. If this function
is not required it should be set to
OFF
by pressing the
^
and
v
buttons simultaneously at this
function.
5.38
High overrange visual warning limit value
Display:
HIGH dISP
Range:
Any display value or
OFF
Default Value:
OFF
High overrange limit value - the display can be set to show an overrange message if the display
value rises above the
HIGH dISP
setting. For example if
HIGH dISP
is set to
1000
then once
the display reading rises above
1000
the message
-or-
will flash on and off or the display value
will flash on and off instead of the normal display units (see
dISP
function
). This message
can be used to alert operators to the presence of an input which is above the high limit. If this
function is not required it should be set to
OFF
by pressing the
^
and
v
buttons simultaneously
at this function.