d.HLd
- display hold mode. The 7 segment display and retransmission value will be held whilst
the remote input display hold switch is closed.
Hi
- peak (max.) memory mode. With the peak remote input switch open the retransmission
will indicate the peak value in memory i.e. the retransmission output can rise but not fall. The
retransmission output can be reset by clearing the memory. The memory may be cleared either
by closing the remote input switch for approximately 2 seconds or by removing power to the
instrument.
Lo
- valley (min.) memory mode. With the valley remote input switch open the retransmission
will indicate the valley (min.) value in memory i.e. the retransmission output can fall but not
rise. The retransmission output can be reset by clearing the memory. The memory may be cleared
either by closing the remote input switch for approximately 2 seconds or by removing power to the
instrument.
dISP
- display mode. The retransmission output will follow whatever value is on the 7 segment
display. For example with the remote input is set to
dISP
the retransmission will follow the rate
when the display is toggled to show rate and will follow the total when the display is toggled to
show total.
5.52
Digital output operation mode
Display:
dG.OP
Range:
rAtE
or
totL
Default Value:
rAtE
Allows the choice of rate or total to be used for the optional digital retransmission output.
5.53
Low overrange visual warning limit value
Display:
Lo dISP
Range:
Any display value or
OFF
Default Value:
OFF
Low overrange limit value - the display can be set to show an overrange message if the display
value falls below the
Lo dISP
setting. For example if
Lo dISP
is set to
50
then once the
display reading falls below
50
the message
-or-
will flash on and off or the display value will
flash on and off instead of the normal display units (see
dISP
function
). This message can
be used to alert operators to the presence of an input which is below the low limit. If this function
is not required it should be set to
OFF
by pressing the
^
and
v
buttons simultaneously at this
function.
5.54
High overrange visual warning limit value
Display:
HIGH dISP
Range:
Any display value or
OFF
Default Value:
OFF
High overrange limit value - the display can be set to show an overrange message if the display
value rises above the
HIGH dISP
setting. For example if
HIGH dISP
is set to
1000
then once
36 of