5.58
PIN code 2
Section:
ACCES
Display:
USR.2 Pin
Range:
0
to
65535
Default Value:
0
Default Access Level
S.CAL
Function number
0C0A
This function allows choice of a second PIN code to be used for PIN code input access method. Associated with
the PIN is an access level (see
Pin.2 AccS
). The second PIN would normally be used to allow a second person
to have a higher access to setup functions via a different PIN. If a second PIN is not required leave the setting at
0
.
If a PIN other than 0 is chosen then this PIN must be entered to gain access to the the selected level.
To access setup functions using the PIN code input access method press then release the
F
button then within 2
seconds press the
^
and
v
buttons at the same time. The message
FUNC
is seen followed by the message
CodE
.
If the message
FUNC End
is seen at this point it means that the access level has been set to
NonE
. Use the
^
and
v
buttons to enter the PIN then press
F
to accept the PIN and proceed to the setup functions. Ony one
CodE
message will appear even though there can be a second PIN. If the number entered into the
Code
at this point is
the PIN code 1 number then access will be granted to the functions allocated to the first PIN. If the PIN code 2 value
is entered then access will be granted to the functions allocated to the second PIN.
5.59
PIN code 2 access level
Section:
ACCES
Display:
USR.2 LEUL
Range:
NONE
,
1
,
2
,
3
,
4
,
5
,
6
,
CAL
Default Value:
NONE
Default Access Level
S.CAL
Function number
0C03
This function allows choice of the access level available when using the PIN code 2 input access method. To access
setup functions using the PIN code 2 input access method press and hold the
F
button until the message
FUNC
is
seen followed by the first function message, this should take approximately 3 seconds. If the message
FUNC End
is seen at this point it means that the access level has been set to
NonE
.
5.60
User assignable access 1 function number
Section:
ACCES
Display:
Fn.1 CodE
Range:
0000
to
FFFF
hex.
Default Value:
0000
Default Access Level
S.CAL
Function number
0C10
In addition to being assigned an access level each setup function is assigned an individual function number. This
functions and the ones which follow (
Fn.2CodE
etc.) can be used to alter the access level for particular functions.
For example if the user wishes to change the access level of the channel 1 display units (function number 43A0)
from level 5 to level 1 then the value
43A0
would be entered at this function and the value
3
would be entered
at the function which follows. This would then enable the channel 1 display unit functions to be accessed at the
lowest access level.
50 of