Instrument setting for absorbance / transmittance measurement
20
①
Set the instrument in the same way as measurement using cuvette.
②
Warming up the light source in the same way as measurement using cuvette.
③
Open the SpectraSmart software, and then open the shutter, slide the side of the
cuvette holder and set the reference film or make air as reference spectrum
(
※
)
.
④
Close the shutter and acquire the dark spectrum
(
※
)
.
⑤
Open the shutter, set the thin film to be measured, and perform the measurement.
※
For the acquisition method, please refer to SpectraSmart User's Guide.
Example: 3. Using probe
Since immersion probe is measured by immersing the probe directly in the sample,
troublesome sampling is unnecessary. 2, 5, 10, 20 mm optical path length tip are
available.
Configuration example
PC (sold separately)
SEC2021 Spectrometer
SEC2022 Deuterium halogen light source
5.4mm Monocoil all legs, Low Sol 300um
UV, 1.5m Total length with round nut
PD-U10 10mm tip
Figure 4-3: Side of the cuvette holder
Figure 4-4: Absorbance/Transmittance measurement example using transmission dip
probe