Data interface
Goldeye Technical Manual
V2.2.0
45
The camera can be single triggered or periodically triggered with a fixed frame
rate (t
p
) up to the maximum frame rate of 100 Hz (Goldeye xy-008) or 30 Hz
(Goldeye xy-032).
In the Goldeye P-... variants the CC1 signal is directly controlled using GigE
Vision features located in the AcquisitionControl branch of the feature tree.
Model
Exposure delay tv
Jitter
Goldeye xy-008
32 μs
(also for reduced resolution)
max. 1 line
Goldeye xy-032
63.4 μs
max. 1 line
Table 19: Exposure delay and jitter
Goldeye xy-008
Goldeye xy-032
t
F
= 10 ms (5.4 ms)
t
p
t
F
(The period duration t
p
must be at least as
long as the image time t
F
)
t
p
– t
exp
> 0.5 ms (Must be kept!)
t
F
= 33.3 ms
t
p
t
F
(The period duration t
p
must be at least as
long as the image time t
F
)
t
p
– t
exp
> 1 ms (Must be kept!)
Table 20: Trigger input diagram
Feature
Description
AcquisitionMode
This feature controls the acquisition mode of the software. This
feature works independently (!) of the chosen camera mode
(Continuous, IOD hardware trigger, IOD hardware timer). It
describes how many frames should be acquired.
AcquisitionStart
Starts the image acquisition of the camera.
AcquisitionStop
Stops the image acquisition of the camera.
Table 21: Camera standard feature: AcquisitionControl