Command reference
R&S
®
ZND
922
User Manual 1173.9557.02 ─ 61
Table 7-11: Set command parameters
One port standards
Two port standards
Open
(Offset)
Short
Match
Reflect
Sliding
match
Symmetric
network
Through,
Line
Attenuation
<Con-
nType>
mandatory
<CalKit-
Name>
mandatory (and must not be an empty string)
<Standar-
dLabel>
mandatory (but can be an empty string)
<MinFreq>,
<MaxFreq>
mandatory
<Delay-
Param>,
<Loss>,
<Z0>
mandatory
mandatory
mandatory
mandatory
not used
mandatory
not used
<C0>,
<C1>,
<C2>, <C3>
mandatory
mandatory
*
mandatory
*
mandatory
**
mandatory
**
not used
<L0>, <L1>,
<L2>, <L3>
mandatory
*
mandatory
mandatory
*
mandatory
**
mandatory
**
OPEN |
SHORt |
MATCh |
<Resist-
ance>
optional
***
optional
***
optional
***
mandatory:
OPEN |
SHORt |
<Resist-
ance>
mandatory:
OPEN |
SHORt |
<Resist-
ance>
<Port1>[,
<Port2>]
optional: <Port 1>
optional: <Port 1>,<Port 2>
*
values are ignored during calibration
**
if OPEN is selected the residual inductance (L0,...,L3) are during calibration; if SHORt is selected the fringing capacitance
(C0,...,C3) are during calibration
***
must be provided if the definition is restricted to a single port. Values are ignored during calibration
The parameters in the
[SENSe<Ch>:]CORRection:CKIT:<StandardType>
,
[SENSe<Ch>:]CORRection:CKIT:<StandardType>:WLABels
, and
[SENSe<Ch>:]CORRection:CKIT:<ConnectorType>:<StandardType>
com-
mands have the following meaning:
Table 7-12: Parameter list
Parameter
Meaning
Comment/Unit
'<ConnType>'
Name of the connector type.
Use
SENSe1:CORRection:CONNection:CATalog?
to query connec-
tor names.
String parameter
'<CalKitName>'
Name of the calibration kit. Use
[SENSe:]CORRection:CKIT:CATalog?
to query cal kit names.
String parameter