Command reference
R&S
®
ZND
901
User Manual 1173.9557.02 ─ 61
<RecallSetFile>
String parameter to specify the absolute or relative path of the
created recall set file. Relative paths are evaluated relative to
the current directory (see
The default extension for recall set files is
znx
. Unless you
specify
znxml
as file name extension, recall sets are always
stored in
znx
file format.
Example:
Usage:
Setting only
Manual operation:
MMEMory:STORe:TRACe
<TraceName>, <TraceFile>[, <FormatInd>[, <Format>[,
<DecSeparator>[, <FieldSeparator>]]]]
Stores the trace data of a specified data trace to a trace file. Traces are created using
the
CALCulate<Ch>:PARameter:SDEFine
Tip:
*.s<n>p
Touchstone files (<n> = 1, 2, 3, ...) are intended for a complete set of
<n>-port S-parameters. Data export fails if the active channel does not contain the full
set of <n>
2
traces or if the involved ports are not numbered consecutively, starting with
port 1. If the necessary traces are available,
'<trc_name>'
can be the name of any
of the traces. To create Touchstone files while less than <n>
2
single-ended traces are
available, use
Setting parameters:
<TraceName>
Name of an existing data trace in the active recall set (string
parameter).
<TraceFile>
String parameter to specify the name and directory of the cre-
ated trace file. Several file formats for trace files are supported.
The file extensions
*.s<n>p
,
*.csv
, and
*.dat
for Touch-
stone, ASCII, and Matlab files are mandatory. To generate a
multiport Touchstone file
*.s2p
,
*.s3p
..., the channel must
contain traces for the full set of S-parameters; the '
<trc_name>
'
is ignored. If no path is specified the analyzer uses the current
directory, to be queried with
MMEMory:CDIRectory?
.
<FormatInd>
FORMatted | UNFormatted
UNFormatted - unformatted data export specified by the second
optional parameter.
FORMatted - formatted data export (for
*.csv
and
*.dat
files
only).
If the first optional parameter is omitted, the command stores
unformatted data.
<Format>
COMPlex | LINPhase | LOGPhase
COMPlex - complex values (real and imaginary part)
LINPhase - linear magnitude and phase.
LOGPhase - dB-magnitude and phase.
SCPI command reference
www.allice.de
Allice Messtechnik GmbH