Command reference
R&S
®
ZND
707
User Manual 1173.9557.02 ─ 61
Parameters:
<SearchFormat>
MLINear | MLOGarithmic | PHASe | UPHase | REAL |
IMAGinary | SWR | DEFault
Identifies the search format for the target value of the marker.
See table above.
*RST:
DEFault
Example:
Suppose channel 1's selected trace is POLar and marker 1 isn't
yet created
:CALCULATE1:MARKER1 ON
Create/enable Marker 1
:CALCulate1:MARKer1:FUNCtion:SELect TARGet
Select TARGet search mode for marker 1
:CALCulate1:MARKer1:SEARch:FORMat?
Query the target format of marker 1. The result is DEF and for
polar diagrams the default target format is "Phase".
:CALCulate1:MARKer1:FUNCtion:TARGet?
Query for the default target value; for "Phase" this is 0 (degrees)
:CALCulate1:MARKer1:SEARch:FORMat MLOGarithmic
Change the target search format to dB magnitude
:CALCulate1:MARKer1:FUNCtion:TARGet?
Query for the default target value; for dB magnitude this is 0 (dB)
:CALCulate1:MARKer1:FUNCtion:TARGet -3
Set the target value to -3 dB
:CALCulate1:MARKer1:FUNCtion:EXECute
Execute the target search for marker 1
:CALCulate1:MARKer1:FUNCtion:RESult?
Query for the results.
Manual operation:
CALCulate<Chn>:MARKer<Mk>:SEARch:TRACking
<Boolean>
This command is only available if a search mode is active (
...
)
It enables or disables tracking, which causes the search logic of the related marker(s)
to be repeated after each sweep.
Tip:
If the current search mode is a bandfilter or multiple peak search this command
enables or disables the corresponding tracking.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Number of an existing marker.
If a bandfilter search or a multiple peak search is active, tracking
recalculates the whole marker set for each sweep and hence
this suffix is ignored.
SCPI command reference
www.allice.de
Allice Messtechnik GmbH