GUI reference
R&S
®
ZND
491
User Manual 1173.9557.02 ─ 61
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:C
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:L
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:R
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:G
Select File...
"Select File..." is enabled as long as the "1-Port Data" network is selected. This net-
work is defined by its S-parameters stored in a one-port Touchstone file (
*.s1p
). No
additional parameters are required.
Note:
The loaded file is stored in the active recall set. Persisted recall sets contain the
full (de-)embedding data so that they can be transferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:GLOop:DEEMbedding
MMEMory:LOAD:VNETworks<Ch>:GLOop:EMBedding<group>
Set to Ideal Ground
This function is enabled as long as the 1-Port Data network is active. An imported S-
parameter set is replaced by the S-parameters of an ideal through connection, which
eliminates the transformation network.
5.13.9
Differential Match tab
Allows you to specify a 2-port embedding network for the differential mode of a bal-
anced port.
Such a network is either defined
●
via its S-parameter stored in a two-port Touchstone file (
*.s2p
) or
●
by specifying the parameters of a "Shunt L, Shunt C" lumped element model
Use the complementary dock widget to access the balanced port configuration and to
activate or deactivate "Differential Match" embedding for selected balanced ports (see
Chapter 5.13.1.8, "Differential Match panel"
Background information
Chapter 4.6.2.8, "Differential match embedding"
Offset Embed softtool
www.allice.de
Allice Messtechnik GmbH