Command reference
R&S
®
ZND
1089
User Manual 1173.9557.02 ─ 61
Example:
SYSTem:TSLock SCReen
Lock the entire screen.
Manual operation:
See
"Enabled / Lock Diagrams / Lock Screen"
SYSTem:TTLout<Pt>:STATus[:STATe]
<Boolean>
CALCulate<Chn>:LIMit:TTLout<Pt>[:STATe]
for
new traces.
Suffix:
<Pt>
.
1 - TTL out pass 1 (pin 13 of USER PORT connector)
2 - TTL out pass 2 (pin 14 of USER PORT connector)
Parameters:
<Boolean>
OFF (0)
New traces are created with
CALCulate<Chn>:LIMit:TTLout<Pt>[:STATe]
set to
OFF
(0)
.
ON (1)
New traces are created with
CALCulate<Chn>:LIMit:TTLout<Pt>[:STATe]
set to
ON
(1)
.
*RST:
Not affected by *RST; factory default is OFF
Manual operation:
See
SYSTem:USER:DISPlay:TITLe
<String>
Defines a title for the remote display, i.e. for the screen that is shown at the instrument,
if a remote session is established and the standard display is switched off.
Parameters:
<String>
Title string.
A \n in the string starts a new line; see example.
*RST:
empty string
Example:
SYST:USER:DISP:TITL "DON'T TOUCH\nRemote test
running..."
Define a title for the remote display.
SYSTem:USER:KEY
<Key>[, <Label>]
Labels a user-defined key in the remote display. In the query form the command
returns whether or not a user-defined key was tapped or clicked.
SCPI command reference