Command reference
R&S
®
ZND
1068
User Manual 1173.9557.02 ─ 61
<Driver>
Matrix driver name indicating the matrix type (string parameter).
Standard drivers:
The supported drivers are displayed in the
"Driver" dropdown-list of the "Add External Switchmatrix" dialog.
Each list entry corresponds to a matrix driver file (
*.matrix
)
located in the
Resources\ExtDev
subdirectory of the analyz-
er's program directory. Specify the <Driver> as the driver's file
name without the
*.matrix
extension, e.g. as
"ZV-Z81-05"
.
Non-standard driver location:
You may also specify the full
absolute path to an appropriate driver file, e.g.
"C:\Program
Files\Rohde-Schwarz\Vector Network
Analyzer\ZNB\Resources\ExtDev\ZV-Z81-05.matrix"
.
<Interface>
Interface type (string parameter)
'USB'
USB interface
'LAN'
LAN interface
<Address>
Interface-specific address (string parameter).
Example:
:SYST:COMM:RDEV:SMATrix1:DEF '', 'ZN-Z84-42',
'USB', '101142'
Register switch matrix no.1: type R&S ZN-Z84 (2x24) with serial
number
101142
at USB interface.
:SYST:COMM:RDEV:SMATrix2:DEF '', 'ZV-Z81-05',
'LAN', '10.10.10.10'
Register switch matrix no.2: type R&S ZV-Z81 (model .05) at IP
address 10.10.10.10.
Manual operation:
See
SYSTem:COMMunicate:RDEVice:SMATrix:DELete
Unregisters all switch matrices, i.e. removes them from the list of configured devices.
Example:
See
SYSTem:COMMunicate:RDEVice:SMATrix:
Usage:
Event
Manual operation:
See
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:RELays:SWITch:COUNt?
For mechanical matrices, this command allows to query the list of switch counts for all
relays (if supported by the matrix).
Suffix:
<Matr>
.
Index of the switch matrix (see
Usage:
Query only
SCPI command reference
www.allice.de
Allice Messtechnik GmbH