Command reference
R&S
®
ZND
1064
User Manual 1173.9557.02 ─ 61
Example:
:INSTrument:SMATrix OFF
Remove all switch matrices from the RF configuration.
:SYSTem:COMMunicate:RDEVice:SMATrix:DELete
Unconfigure all switch matrices, i.e. remove them from the list of
configured devices.
:SYSTem:COMMunicate:RDEVice:SMATrix1:DEFine '',
'ZN-Z84-02', 'LAN', '192.168.0.42'
Register a single switch matrix of type ZV-Z82-30 (with two 2x5
submatrices).
:INSTrument:SMATrix ON
Add the matrix to the RF configuration, performing a default
assignment of VNA ports and test ports.
:SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
STARt
Start a manual RF connection configuration.
:SYSTem:COMMunicate:RDEVice:SMATrix1:CONFigure:
MVNA 1,2,2,1
Define non-default matrix-VNA port connections: matrix port 1 to
VNA port 2, matrix port 2 to VNA port 1 (default is
1,1,2,2
).
:SYSTem:COMMunicate:RDEVice:SMATrix1:CONFigure:
MTESt 1,2,2,3,3,4,4,5,5,6,6,1
Configure 6 matrix test ports: matrix test port i is test port i+1
(modulus 6).
:SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
END
Apply the manual RF connection configuration.
Usage:
Event
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:TVNA
<TestPort1>[,
<VNAPort1>]...
Sets/gets the test port connections of the VNA in a switch matrix RF connection setup,
i.e. the direct assignments of test ports to physical VNA ports.
Note:
a manual RF connection configuration must be performed within a transaction,
enclosed by the commands
SYSTem:COMMunicate:RDEVice:SMATrix:
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
. The modified configuration will not be applied until the end of this transaction.
Setting parameters:
<TestPort1>
Number of the test port.
Test ports must be numbered subsequently starting at 1.
<VNAPort1>
Number of the VNA port
Example:
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
TVNA 7,2
Test Ports 7 is assigned to VNA port 2.
Manual operation:
See
SCPI command reference