Cycle operating
Preparation
Sequence
of
operations
Opération
Symbol
1
Connect the water supply hose,
Connect the drain hose,
Open the water tap
2
Connect the power cable, enable the system's circuit
breaker,
3
Make sure the drain plug is securely in place
4
Fill the hot-water bath with 2.5L of water (Water level just
under the 2
nd
turn of the heating component),
5
Put the bucket into the vessel, a bucket lock orientated at
10
Go the bucket down to the bottom and turn in the clockwise
direction until It is locked in the vessel.
PAY ATTENTION NOT TO HIT THE TEMPERATURE PROBE
Clip
6
Place the stirrer on the axis of the
bucket after après s’être
assurer de la présence de la rondelle de glisse sur l’axe,
7
Replace the carded cotton in the cage (Before each cycle)
8
Switch on the device by pressing the main switch
9
Pour 1/3 of demineralized water of the total volume to be
prepared in the vessel
10
Slowly pour the powder, avoiding putting it on the wall and
on the blades of the stirrer
11
Start the agitator at speed 5