37
Navigating and Customizing Your FP-25
Sensor Setup
The SENSOR SETUP menu contains advanced settings that govern how the flow
and pressure sensors report their data.
PRESS AVG changes the length of the geometric running average for pressure
readings. Values roughly correspond to the time constant (in milliseconds) of
the averaged values. Higher numbers generate a greater smoothing effect on
quickly fluctuating readings (default is 255, full averaging).
ZERO BAND is the deadband threshold, a value below which the FP-25 displays
all flow readings as 0 (no flow). This function also applies to gauge pressure
readings from the internal pressure sensor. The FP-25 ships with a ZERO BAND
value of 0.0%, which means that all flow values are reported, no matter how
tiny. Increasing this value effectively chops off the bottom end of the device's
flow range. This can be useful if you want to disregard all flow rates below the
turndown range of the instrument , and so they display as zero flow.
Note:
ZERO BAND
values do not affect the serial data frame.
Advanced Setup |
Sensor Setup
MAIN
BACK
PRESS
AVG
ZERO
BAND
MAIN Returns to the Instant
screen of Standard Mode or
Basic Mode, whichever was
last used.
BACK Returns to the Advanced
Setup Menu (page 36).
PRESS AVG Adjusts the length
of the geometric running
average (1-255 ms) for
reported pressure values.
ZERO BAND Sets the deadband
threshold (max 6.3% of full
scale), under which flow
values are displayed as 0.
001
Avg Time Const: msec
SET
CLEAR
BACK/
CANCEL
SELECT
DIGIT
DOWN
UP
>
0.0
Zero: % of Full Scale
SET
CLEAR
BACK/
CANCEL
SELECT
DIGIT
DOWN
UP
>