background image

 

 

 

Initial Installation 

The RX and TX bars should be placed on either side of a treadmill or in a large, open floorspace with the LED 
strips facing each other. When setting up a treadmill gait test, the interface sides of the bars should be placed 
at the front of the treadmill.  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1)

 

After placing the TX and RX Bars in the correct position, plug in the power cables for each bar. When an 
OptoGait system is being used for the first time, the RX and TX bars may need to be charged prior to use.  

 

2)

 

The Lemo (if EMG is in use) and USB cables are plugged into the RX bars and then directly into the lap 
top/computer via a USB port. Please note:  A standard OptoGait set-up requires three computer USB port 
to connect the RX bar cables and Logitech Cameras.  

 

3)

 

The Cameras should be placed on a side and rear view (see 

Necessary Equipment

 for reference). When 

using the OptoGait system for research purposes, a frontal view camera may be adopted.  

 

4)

 

The OptoGait system can be switched on using the power switches, located on the interface side of both 
the TX and RX bars.  

 

5)

 

Initially, a red LED light will appear and then change to green after 2-3 seconds to show that the TX and 
RX bars are connected to each other. In the case that no green

 

LED light appears, make sure that the 

bars are parallel to each other and that no external objects are obstructing the floor space/treadmill belt.  

 

6)

 

After the RX and TX bars are connected to each other, switch on the computer/laptop and open up the 
OptoGait software. The OptoGait loading screen then appear and typically take 1 – 2 minutes to load.  

Interface Side of Bars 

LED strips 

Summary of Contents for OptoGait

Page 1: ...OptoGait User Guide Installation and Use...

Page 2: ...ptoGait Software 3 28 Camera Setup 3 4 Hardware Test 5 6 Additional Devices 7 8 Creating a Patient 9 10 Setting Up a Test 11 14 Running a Test 15 18 Managing Data 19 22 Using Biofeedback Mode 23 25 Su...

Page 3: ...ptoGait system Lemo cable small silver cable with red dot x1 Optional used In combination with EMG Power cables for charging the TX and RX bars x2 USB cable Laptop charging cable for extensive data co...

Page 4: ...ct the RX bar cables and Logitech Cameras 3 The Cameras should be placed on a side and rear view see Necessary Equipment for reference When using the OptoGait system for research purposes a frontal vi...

Page 5: ...n Utilities in the top right hand corner 2 Go to the Video section and select Video 1 and Video 2 for example Logitech Pro Webcam C920 Logitech Pro Webcam C920 2 3 Click on Test and select the gain an...

Page 6: ...the camera zoom contrast brightness saturation sharpness and focus This section allows the user to select the optimum image quality for all aspects of the video image 5 To finalise any changes made to...

Page 7: ...5 Hardware Test 1 To perform a OptoGait hardware test click on Utilities in the top right hand corner 2 Click on OptoGait hw Test 3 To execute a hardware test select Execute...

Page 8: ...standard hardware test Any interferences between the LED s will appear in red on the diagram below the video image If there are no interferences between the LED s then testing can begin 5 If there ar...

Page 9: ...see all the available devices If the additional device has been connected correctly the device will appear in the Available Devices section 4 After selecting the device click on the specific device an...

Page 10: ...8...

Page 11: ...r On this page the patients name date of birth height weight gender an image for reference purposes and other information can be added 4 The patients foot size can then be measured by clicking Acquire...

Page 12: ...the three ways to save the patient s profile all located in the top right hand corner includes Save Save and go back to the OptoGait home screen Save and New Save and create a further new patient Save...

Page 13: ...he clinician to select the patient as well as the type of exercise and protocol they wish to run 3 To create a test click on Define Modify Tests 4 Click on New Test 5 Input the test details including...

Page 14: ...Duplicate Test or transferred by clicking on Export Test or Import Test 8 An existing test for example Treadmill Walking 3MPH or 5KP H can be selected from the scroll down list labelled Test 9 To cre...

Page 15: ...fter clicking New Protocol Select the protocol s and click on the sideward arrow to move them to the section labelled Protocol 12 Click Save 13 Protocols can be modified or deleted by clicking on Edit...

Page 16: ...14 14 An existing protocol Body Sway Protocol Single Leg 3 Hops 2D Protocol etc can be selected from the scroll down list labelled Protocol...

Page 17: ...the patient test and protocol 2 If the software visuals need to be altered click on Configuration and choose to either show or hide charts display video and quantitative data and alter the overall exe...

Page 18: ...pop up will appear highlighting the direction of the treadmill belt and selected speed After selecting the speed on the OptoGait software set the same speed on the treadmill 5 To start the test click...

Page 19: ...foot After clicking on the left or right foot icon data collection will begin 7 If the test is being stopped by a Software Command manually click Stop to finish the test Otherwise if the test is bein...

Page 20: ...splay video footage from both cameras and a pie chart demonstrating the time that the patient spent within pre selected thresholds for the variable of choice for example the figure below shows that th...

Page 21: ...d click on the sideward arrow to move the data from the Test List left to the Test Analysis right 3 Click View to look at the results screen On this screen two pieces of data e g returning visits foll...

Page 22: ...and time parameters can be viewed by using the scroll bar in the Gait Report section 5 In the Gait Report section click on the magnifying glass icons displayed in green or yellow to view the muscle ac...

Page 23: ...rint to print a data report There is also the option to add markers to specific parts of the video file by clicking Markers 7 The video files from both cameras can be played and skipped in the top lef...

Page 24: ...22 8 The activation of each lower extremity muscle during gait can be viewed here by using the scroll bars...

Page 25: ...g Biofeedback 1 To access Biofeedback mode click on Biofeedback on the top tool bar 2 To start a test click one of the following test options Treadmill walking test Treadmill running test March in pla...

Page 26: ...s by clicking on the three horizontal line icon in the top right hand corner 4 The Opto for Kids can be selected as a more engaging method of gait analysis for children For this option the child must...

Page 27: ...feedback test a pie chart will be displayed showing the time and percentage that the customer spent with the good green fair orange and red bad thresholds for the selected gait parameter The threshold...

Page 28: ...tandard phases of gait click on GMF Gait Muscles and Functions on the top tool bar 2 Click on the drop down arrow to view a particular phase of the gait 3 To alter the speed play or skip the video use...

Page 29: ...27...

Page 30: ...rther Technical Support 1 To be redirected to the full OptoGait manual click on Help on the top tool bar For further technical support from Algeos contact Ben on 44 0 151 448 1228 or on support algeos...

Reviews: