PDA Mobile Software User's Guide
61
22 Data Transfer to PDA via IR
This section describes how to transfer cell data from a Cellcorder and
specific gravity (SG) / temperature readings from a hydrometer to the PDA
via the IR (infrared) interface.
22.1 File Transfer from Cellcorder to PDA
Before transferring data to the PDA, load the data in Cellcorder temporary
memory, press the Cellcorder Shift+Send, then make menu selections. On
the PDA Main Page, press Cellcorder, follow screen instructions, then press
Download to transfer.
Figure 34. Download Cellcorder
The Cellcorder displays
Sending data for Cell
nnn
during transfer. Upon
completion, the PDA displays
Cellcorder Download Complete
. Press OK to
store the file. On the Cellcorder, return to the Main Menu. If you need to edit
location, battery or string names, refer to
Site Information
.
22.2 Data Transfer from Hydrometer to PDA
To transfer hydrometer module SG and temperature readings to the PDA, on
the module, select a string A to H. On the PDA Main Page, press
Hydrometer, follow screen instructions, then press Download to transfer.
The hydrometer module LEDs go off during transfer. Upon completion, the
PDA displays
Hydrometer Download Complete
. You may edit location,
battery or string names on this page, then press OK to store the file.
To edit location, battery or string names:
Press the stylus on the text or
number to be edited, then press the ABC or 123 button on the PDA. Edit the
text or number, then press Done.