AirSep
®
Corporation
12/02
AS-20—1000 Instruction Manual
7-3
7.4
Depressurizing the Filters
The oxygen generator filters contain pressurized air. To
perform monthly maintenance or change filter elements safely,
fully depressurize the filter assembly as described in the
procedure in this section. To safely depressurize all oxygen
generator components, refer to the procedure described later in
this chapter. (See Section 7.6, Depressurizing the Oxygen
Generator.)
1) Make sure the AUTO/MAN switch is set to
AUTO
and allow the oxygen
generator to stop cycling when the oxygen receiver pressurizes fully.
2)
Disconnect the oxygen generator from the power supply.
3) Remove the front cover of the enclosure, if applicable. (See Section 7.3,
Removing the Front Cover of the Enclosure.)
4)
Fully close the manual inlet valve and manual filter outlet valve. (See Figure
7-1.)
Coalescing Filter
Particulate filter
(not provided on
AS-20 model)
Feed Air Regulator
Filter Bowls
Bowl Latches
Manual Filter
Outlet Valve
Drain Tubing
Connections
Exhaust Muffler
Manual Inlet Valve
Pressure Switch
Figure 7-1 Interior View
of AS-45 and AS-80 Models
Summary of Contents for AS-1000
Page 2: ......
Page 4: ......
Page 14: ......
Page 16: ......
Page 24: ......
Page 32: ......
Page 38: ......
Page 48: ......
Page 61: ...AirSep Corporation 12 02 AS 20 1000 Instruction Manual A 9 Operating Data ...
Page 62: ...12 02 AirSep Corporation A 10 AS 20 1000 Instruction Manual Drawings ...
Page 64: ...12 02 AirSep Corporation A 12 AS 20 1000 Instruction Manual Figure 3 Flow Schematic AS 20 ...
Page 78: ...12 02 AirSep Corporation Figure 21 Flow Diagram AS 1000 A 26 AS 20 1000 Instruction Manual ...
Page 82: ......