AS-20—1000 Instruction Manual
7-7
Use extreme caution when performing this procedure since it
requires removal of the front cover of the enclosure while the
unit receives electrical power.
Refer to Appendix D of this instruction manual for additional
information on the pressure switch.
1)
Set the ON/OFF switch on the oxygen generator control panel to ON.
2)
Open the front cover of the enclosure and locate the pressure switch. (See
Figure 7-1.)
Locate the two adjustment screws on top of the pressure switch. The
adjustment screw closest to the rear of the cabinet controls the high pressure
setting. The adjustment screw closest to the front of the cabinet controls the
low pressure setting. Use a ¼-inch wrench or screwdriver to make
adjustments.
The adjustment screw normally requires less than a ¼ turn.
Turn either the high or low adjustment screw ¼-turn clockwise
to increase its pressure setting. Turn either screw
counterclockwise to decrease its pressure setting.
To prevent damage to the pressure switch, do not force the
adjustment screws past their stopping points.
3)
Set the AUTO/MAN switch on the oxygen generator control panel to
AUTO
.
Summary of Contents for AS-1000
Page 2: ......
Page 4: ......
Page 14: ......
Page 16: ......
Page 24: ......
Page 32: ......
Page 38: ......
Page 48: ......
Page 61: ...AirSep Corporation 12 02 AS 20 1000 Instruction Manual A 9 Operating Data ...
Page 62: ...12 02 AirSep Corporation A 10 AS 20 1000 Instruction Manual Drawings ...
Page 64: ...12 02 AirSep Corporation A 12 AS 20 1000 Instruction Manual Figure 3 Flow Schematic AS 20 ...
Page 78: ...12 02 AirSep Corporation Figure 21 Flow Diagram AS 1000 A 26 AS 20 1000 Instruction Manual ...
Page 82: ......