35
Chapter 4
Maintenance
The Model TA465 and probe accessories require very little
maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model TA465, 960 series
thermoanemometer probes, IAQ and VOC probes to TSI for annual
recalibration. Please contact one of
TSI’s offices or your local
distributor to make service arrangements and to receive a Return
Material Authorization (RMA) number. To fill out an online RMA form,
visit TSI’s website at
TSI Instruments Ltd.
Stirling Road
Cressex Business Park
High Wycombe
Bucks
HP12 3ST United Kingdom
Tel: +44 (0) 149 4 459200
Fax: +44 (0) 149 4 459700
The Model TA465 and accessory probes can also be recalibrated in
the field using the CALIBRATION menu. These field adjustments are
intended to make minor changes in calibration to match a user's
calibration standards. The field adjustment is
NOT
intended as a
complete calibration capability. For complete, multiple-point
calibration and certification, the instrument must be returned to the
factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with
a soft cloth and isopropyl alcohol or a mild detergent. Never immerse
the Model TA465. If the enclosure of the Model TA465 or the AC
adapter becomes broken, it must be replaced immediately to prevent
access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month
to prevent damage due to battery leakage.