background image

Air Techniques, Inc.

Page 20

Figure 10.  

AERO100 and AERO200 Parts Location (AERO100 shown)

1

3

4

2

5

13

7

12

11

14

9

8

8

10

6

4

Appendix 3: Spare Parts 

Summary of Contents for AERO100

Page 1: ...AERO100 AERO100M AERO200 AERO200M OPERATING INSTRUCTIONS AERO200M...

Page 2: ...nical information Qualified personnel This operating manual is valid for oil free compressor stations hereinafter called product The operating manual is only valid so long as your product complies wit...

Page 3: ...trical installation 9 2 3 4 Motor protection 9 2 3 5 Circuit diagram 10 2 4 Initial operation 10 3 Operation 11 3 1 Switching on the product 11 3 2 Switching off the product 11 3 3 Switching on the pr...

Page 4: ...ate authority Operator Authority Responsible for the safe installation regular servicing and cleaning of the product Qualified personnel Trained by the operating authority or by authorised Air Techniq...

Page 5: ...trical components and if they are damaged have them repaired by a qualified electrician Check the electrical equipment for external damage before work starts Check thoroughly whether lines or cables a...

Page 6: ...ompressed air until the set shut off pressure of 7 bar is indicated in the pressure switch 3 before switching off The tank pressure is indicated by the pressure gauge 9 Pressure is re leased in the pr...

Page 7: ...the membrane fibres thereby the water molecules contained in the air diffuse through the membrane wall and accumulate outside of the fibre The dried air flows now via the fine filter the relief valve...

Page 8: ...oduct must only be installed and operated in dry well ventilated and dust free rooms Take care when choosing the site for installation that the product is easily accessible for operating cleaning and...

Page 9: ...ng system If the equipment is hard wired to the power supply a switch arrangement must be arranged with at least a 3 mm wide contact opening e g heavy duty switch If the product is connected to the po...

Page 10: ...re switch is off Check the connections are correct to the compressed air components Make sure the machine is correctly connected to the power supply Check that the air filter is correctly fitted Switc...

Page 11: ...duct Switch off the product to setting 0 OFF 3 3 Switching on the product after a power supply interruption Advice The product will not start against pressure Therefore the products are fitted either...

Page 12: ...pressure reducer By slightly lifting the regulator knob 1 the constant pressure can be altered Increase pressure by rotating the regulator knob clockwise or decrease pressure by turning the regulator...

Page 13: ...me working conditions e g long work times under full load high ambient temperature very dusty atmosphere high humidity the maintenance intervals will need to be shorter When carrying out maintenance w...

Page 14: ...unterclockwise until the bayonet lock is closed again Attention Never clean filter inserts with petrol or oil 4 2 3 Check safety valve function The safety valve see Figure 7 is set in the factory at t...

Page 15: ...are parts are not available the compressor station must be sent to the manufacturer for repair Figure 9 Change of Fine or Sterile Filter Note Membrane Dryer is shipped with a fine filter A sterile fil...

Page 16: ...are exposed to pressure It must be ensured that the internal pressure will not permanently exceed the maxi mum working pressure PS stipulated on the vessel nameplate however it may momentarily do so...

Page 17: ...and dimensions Specifications AirStar AERO100 100M AirStar AERO200 200M Dimensions H x W x D cm 68 x 64 x 49 68 x 64 x 49 Weight kgs 43 54 Delivery rate at Pe 0 bar 110 l min 0 bar 195 l min Delivery...

Page 18: ...reater surface area Gasket defective Replace Cup seal leaking Remedy 1 Worn or pitted 1 Exchange cup seal cylinder and washers spare part kit 2 Very dirty 2 Change filter 3 Too high ambient temperatur...

Page 19: ...7619 87620 9 Safety Relief Valve 87621 10 Tank Drain Valve 87622 11 Starting Sphere C1188 AERO100 and AERO200 ONLY 12 Non return valve receiver 87623 AERO100M and AERO200M ONLY 12 Non return valve rec...

Page 20: ...Air Techniques Inc Page 20 Figure 10 AERO100 and AERO200 Parts Location AERO100 shown 1 3 4 2 5 13 7 12 11 14 9 8 8 10 6 4 Appendix 3 Spare Parts...

Page 21: ...Note Membrane Dryer is shipped with a fine filter 17 A sterile filter 18 is an option Figure 11 AERO100M and AERO200M Parts Location AERO200M shown 3 2 13 8 16 15 5 6 5 6 9 12 3 2 10 4 14 7 Internal...

Page 22: ...er and article number Order number as per parts list How many of each item required Exact address to dispatch to Dispatch details Repairs return delivery Please follow the packaging requirements liste...

Page 23: ...Page 23 Air Techniques Inc Notes...

Page 24: ...their practices to the next level Air Techniques family of quality products for the dental professional include Corporate Headquarters 1295 Walt Whitman Road Melville New York 11747 3062 Phone 800 24...