3.8 Humidifier
If this ventilator is used with a humidifier, ensure that it is always placed lower than the ventilator and
the patient.
It is also recommended that you use patient circuits equipped with water traps when using a humidifi-
er.
CAUTION:
Remember to empty the water traps regularly during ventilation.
CAUTION:
Ensure that water does not enter the unit during handling of the patient circuit
or the humidifier (if used). If this occurs, immediately stop using the device, and contact the
Technical Department.
CAUTION:
Humidification may increase the resistance of the filters used in the patient cir-
cuit. The filters should be tested frequently to check for an increase in resistance or block-
age.
CAUTION:
If a nebulizer is used, it is necessary to use a filter on the expiratory limb to avoid
damage to the expiratory flow sensor.
3.9 Nebulizer
CAUTION:
Y-piece respiratory filters can prevent medication from being effective: their use is
therefore not recommended,
CAUTION:
The precision of the expired volume can be impaired: a protective filter can then
be used at the expiration end.
CAUTION:
Nebulization may increase the resistance of the filters used in the patient circuit.
The filters should be tested frequently to check for an increase in resistance or blockage.
CAUTION:
If a nebulizer is used, it is necessary to use a filter on the expiratory limb to avoid
damage to the expiratory flow sensor.
3.10 Extractable battery
YYYY / MM
YYYY / MM
/
When using the extractable battery for the first time, indicate the month of expiry on the label (two
years after the commissioning date).