9 Technical description
Adult
Child
Infant
Thresh-
old (name)
Ventilation modes
Set-
ting
by
Min~Max [default]
High Chest
Compression
Frequency (fCC)
CPV
5
110~140 [120]
Not available
Maximum pres-
sure variation
of chest com-
pressions (P-P)
CPV
NA
Not adjustable
Not available
9.2.6 Automatic alarm thresholds
This table shows the calculation of the alarm threshold values when
Auto. set
is pressed.
For further information, see Automatic thresholds on page 58.
Thresh-
old
(name)
(A)VCV
(A)PCV
PSIMV
PSV
VSIMV
CPAP
APRV
PRVC
PS-Pro
CPV
High
pressure
(Ppeak)
Mea-
sured
pressure
+ 33%
Measured pr
max value (20% or 5)
No auto threshold
Measured pr
max value (20% or 5)
Low
pressure
(Ppeak)
Mea-
sured
pressure
- 33%
If low Ppeak higher than 5: Mea-
sured pressure - max value (20%
or 5)
Otherwise, 1
No auto threshold
If low Ppeak higher than 5: Mea-
sured pressure - max value (20%
or 5)
Otherwise, 1
Low res-
piratory
rate (RR)
Measured RR - 50% (limited to 4 c/min minimum)
High res-
piratory
rate (RR)
Measured RR + 50% (limited to 90 c/min maximum)
Low VTe
Measured
No auto
thresh-
old
Mea-
sured
VTe -
75%
Default
values
No auto threshold
Mea-
sured
VTe -
50%
High
VTe
Measured
VTe + 50%
No auto
thresh-
old
Mea-
sured
VTe +
Default
values
No auto threshold
Mea-
sured
VTe +
50%
Low VTi
Measured
VTi - 50%
No auto
thresh-
old
Mea-
sured
VTi -
75%
Default
values
No auto threshold
Mea-
sured
VTi -
50%
High VTi
Measured
No auto
thresh-
old
Mea-
sured
VTi +
Default
values
No auto threshold
Mea-
sured
VTi +
50%
Low
MVe
Auto threshold: Measured MVe - 50%
135