9 Technical description
9.2.5 Alarm threshold settings
Adult
Child
Infant
Thresh-
old (name)
Ventilation modes
Set-
ting
by
Min~Max [default]
Volumetric modes:
(A)VCV
10~80 [45]
Pressure-con-
trolled modes:
(A)PCV, PSV, CPAP,
VSIMV, APRV, PSIMV,
PRVC, PS-Pro
max value (10 or PI + 5)~80 [45]
High pres-
sure (Ppeak)
CPV
1
10~80 [60]
Not available
Volumetric modes:
(A)VCV
OFF then 1~75 [1]
Pressure-con-
trolled modes:
(A)PCV, PSV, CPAP,
VSIMV, APRV, PSIMV,
PRVC, PS-Pro
1
OFF then 1 ~ min value (75 or PI) [1]
Oxygen therapy
Not adjustable
Low pres-
sure (Ppeak)
ECLS activated
1
OFF then 1~75 [OFF]
Not available
All except Hflow
1~50 [6]
1~80 [10]
1~80 [20]
Low respirato-
ry rate (RR)
ECLS activated
1
1~50 [1]
1~80 [1]
NA
High respira-
tory rate (RR)
All except Hflow
1
11~60 [35]
11~100 [40]
11~120 [60]
Pressure-con-
trolled modes:
(A)PCV, PSV, CPAP,
APRV, PSIMV, PRVC,
PS-Pro
OFF then 10 ~ min val-
ue (2,000 or high VTe
threshold - 50) [OFF]
OFF then 10 ~ min val-
ue (1,950 or high VTe
threshold - 50) [OFF]
OFF then 10 ~ min
value (790 or high VTe
threshold - 10) [OFF]
Volumetric modes:
(A)VCV, VSIMV
OFF then 10 ~ min
value (2,000 or
Vt setting) [OFF]
OFF then 10 ~ min
value (1,950 or
Vt setting) [OFF]
OFF then 10 ~ min
value (790 or Vt
setting rounded up
to the next ten) [OFF]
Low exhaled
volume (VTe)
ECLS activated
10
OFF then 10 ~ min
value (2,000 or
Vt setting) [OFF]
OFF ~ min val-
ue (1,950 or Vt
setting) [OFF]
Not available
Pressure-con-
trolled modes:
(A)PCV, PSV, CPAP,
APRV, PSIMV, PRVC,
PS-Pro
max value (50 or
low VTe thr
50) ~ 3,000 [2,000]
max value (50 or
low VTe thr
50) ~ 2,000 [1,000]
max value (20
or Vt s
10) ~ 800 [100]
High exhaled
volume (VTe)
Volumetric modes:
(A)VCV, VSIMV
10
max value (50 or
Vt s 50)
~ 3,000 [2,000]
max value (50 or
Vt s 50)
~ 2,000 [1,000]
max value (20 or
low VTe threshold
+ 10) ~ 800 [100]
133