18 - Specification
Specification Conditions
93
SMU4000 Series Instruction Manual
SPECIFICATION
Specification Conditions
Settling Time and Accuracy specifications are valid after a minimum of 60 minutes warm-up
followed by self-calibration execution and measurement zero. Ambient temperature change of
less than ± 3°C after self-calibration and measurement zero.
Calibration Period:
1 Year
Temperature:
23°C ± 5°C
Measurement Period:
≥10 PLC
Key Specifications of SMU4000 series
SMU4001
SMU4201
Maximum Voltage
±21V
±210V
Voltage ranges
20mV – 20V
20mV – 200V
Operating Modes
Source Voltage (SV), Source Current (SC), Load Current (LC), Load Resistance (LR), Load Power
(LP), Measure Voltage (MV), Measure Current (MC), Measure Resistance (MR), Measure High
Resistance (MHR), Sequence.
Voltage Measurements (all modes except MC)
Range
Resolution
Accuracy
(%r offset)
20mV
10nV
±0.06%rdg ± 50µV
200mV
100nV
±0.015%rdg ± 50µV
2V
1uV
±0.015%rdg ± 50µV
20V
10uV
±0.02%rdg ± 5mV
200V¹
100uV
±0.02%rdg ± 5mV
1
200V range only available on SMU4201
Current Measurements (all modes except MV)
Range
Resolution
Accuracy
(%r offset)
200nA²
100fA
±0.04%rdg ± 300pA
2uA
1pA
±0.03%rdg ± 300pA
20uA
10pA
±0.02%rdg ± 2nA
200uA
100pA
±0.02%rdg ± 2nA
2mA
1nA
±0.02%rdg ± 200nA
20mA
10nA
±0.02%rdg ± 2µA
200mA
100nA
±0.02%rdg ± 20µA
1A
1uA
±0.07%rdg ± 100µA
3A³
1uA
±0.2%rdg ± 300µA
2
Typical Accuracy
3
Only available by Manual range selection when in modes that utilize the current source