8 - Manual setup
Source shape
53
SMU4000 Series Instruction Manual
Pulse
When a Pulse shape is selected, the SMU provides a pulsed source between two set levels with
variable width control of both set levels.
Settings
The effective width of each pulse level is controlled by the pulse level measurement setting. This
is equivalent to the measurement period but is separated into two individual settings, one for
each level. The pulse Levels (1
st
and 2
nd
) can be updated from the Manual Setup screen and the
Home screen. When the output is enabled, “real-time” adjustments can be made from the
Home screen, see ‘
Live adjusting source and limits from the home screen
Timing
All pulse timing is based upon the settling/ measure delay, pulse level measurement period and
measurement count. The measure delay shown in the following diagram shows a user defined
measure delay. Alternatively an automatic settling delay can be selected. See ‘
for more details.