9 - Remote Operation
36
QPX750 Instruction Manual
9.4.5
Status Model
9.5
Remote/ local operation
At power-on, the instrument will be in the local state and the REM indicator is not displayed on the
Status Line. In this state, all front panel operations are possible.
When the instrument receives a command from an interface the remote state will be entered, and the
REM indicator will be displayed on the Status Line. In this state the front panel user interface can still
be navigated but no changes to parameters are possible.
The instrument may be returned to the local state by pressing the LOCAL key; however, the effect of
this action will only remain until the instrument receives another character from the interface, when
the remote state will once again be entered. Returning to Local by this action will keep the settings at
their last remotely set values.
NOTE
This instrument is backwards compatible with the QPX600DP remote command set.
These can be found in the QPX600DP instruction manual, see