11.2.6 Memory Output
The content of the data memory can, com-
pletely or in parts, be output to the serial in-
terface. For each output one of the three
available output formats ´List´, ´Columns´ or
´Table´ can be used. The option to specify
partial ranges is available as it is possible to
set the start and end time of measurements
and also possible to select the number of
corresponding identified measurements.
With
external SD memory cards
(see 10.2.1) there is only one op-
tion available, namely output in ta-
ble mode of all the measured data contained in the file most recently
used. For this purpose simply press the key
in the function
Memory free
in the menu
Memory output
or in certain measuring
menus.
The most sensible approach is to remove the memory card and copy
the files via a USB card reader directly onto the PC. These can then
be imported either into MS-Excel or into Win-Control (as of V.4.9).
Menu
Memory Output
:
Setting the
output format
:
Output form: List
In case of
selecting a numbered measurement:
NUMBER: 12-001 A
In Function
Number
select number with keys:
<FIRST>
,
<NEXT>
...,
<LAST>
To
select a time interval:
Enter the start time using the format ´
hh:mm:ss
´:
Start time:
07:00:00
Enter the end time using the format ´
hh:mm:ss
´:
End time:
17:00:00
Enter the start date using the format ´
dd:mm:yy
´:
Start date:
01.05.00
Enter the end date using the format ´
dd:mm:yy
´:
End date:
01.05.00
Perform complete output of data memory:
<ALL>
Output of a numbered measurement:
<NR>
Output of a time interval from start to end:
<TIME>
Cancelling the memory output using the key:
<STOP>
The memory contents will be output with the same print format as with a printer
operation, including multiple printouts and different formats (s. man. 6.6.1).
During the output of data memory, the remaining memory content (in kB) to be
output is continuously indicated in the function
Residual output
. Time, date
and number show the current values .
Remaining memory output:
Residual output: 12.5 kB
current number of memory output:
Number: 01-001A
current time and date of memory output:
Time: 12:34:56 Date: 01.01.04
ALMEMO
®
2890-9
53
* OUTPUT FROM MEMORY *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Memory Internal:
512.0 kB
Memory free:
125.8 kB
Residual output:
12.5 kB
Output form:
Columns
Number:
01-001 A
Time: 12:34:56 Date: 01.01.04
Time interval:
Start time:
07:00:00
Start date:
01.01.04
End time:
17:00:00
End date:
01.01.04