17. Output modules
Trigger inputs
For the purpose of controlling the measuring se-
quence two trigger inputs are provided at ports
8 and 9 (keypad or optocoupler). On the relay
trigger analog adapter (ZA-8006-RTA3) the trig-
ger source 'key' and / or 'optocoupler' can be
configured by means of keys
PROG
,
▲
/
▼
and
PROG
or the trigger function can, for safety
reasons, be switched off altogether by means of 'off'.
The following trigger functions can be programmed as variants:
0: Start and stop a measuring operation
0: Start / Stop
1: Once-only manual measuring point scan
1: Once-only scan
2: Delete all maximum / minimum values
2: Delete all max./min. val.
3: Print the measured value
3: Print
4: Start / stop a meas. op., level-controlled
4: Start/Stop level-controlled
8: Set measured value to zero
8: Set meas. value to zero
-5: Execute macro 5 (s. Man. 6.6.5)
-5: Macro5
-6: Execute macro 6
-6: Macro6
-7: Execute macro 7
-7: Macro7
-8: Execute macro 8
-8: Macro8
-9: Execute macro 9
-9: Macro9
17.3 Analog outputs
V5 output modules
For the purposes of analog recording of mea-
sured values it is still possible, at sockets A1
and / or A2
(2)
to connect V5 output modules
with an analog output driven by the device, e.g.
recording cable ZA-1601-RK (s. Hb. 5.1.1).
To select the socket press
<P>
:
▲
or
▼
The following output modes can be programmed as variants:
0: Measured value for the selected measuring channel
0: Selected channel M00
2: Measured value for a programmed channel
2: Assigned internally M01
8: Programmed analog output (see below)
8: Driven externally
The analog value appears below this in digits:
Analog value: 08345
This gives, depending on the analog output, the following output signals.
Voltage output
-1.2 to +2.00 V
0.1mV/Digit
Voltage output
-6.0 to +10.0 V
0.5mV/Digit
Current output
0.0 to 20.0 mA
1µA/Digit
In variant 2 'assigned internally', after selecting the Mxx function, you can pro-
gram the measuring point to be output
2: assigned internally M
60 ALMEMO
®
204
* OUTPUT MODULES *
Socket: A2
RC Recording cable
0: selected Channel M00
Analog value: 02234
Mªª MENU P ©P
* OUTPUT MODULES *
Socket: A2 ZA 8006 RTA3
Port: 8 Adr.:28
Trigger: keap.+Optocoupler
0: Start-Stop
Mªª MENU P