Specifications and Characteristics
3
U2000 Series Operating and Service Guide
47
Maximum Power
Models
Maximum Power
U2000/1/2A
+25 dBm (320 mW) average, 20 VDC
+33 dBm (2 W) peak, <10
μ
s
U20004A
+25 dBm (320 mW) average, 5 VDC
+33 dBm (2 W) peak, <10
μ
s
U2000/1H
+33 dBm (2 W) average, 20 VDC
+50 dBm (100 W) peak for 1
μ
s
U2002H
+33 dBm (2 W) average, 10 VDC
+50 dBm (100 W) peak for 1
μ
s
U2000/1B
+45 dBm (30 W) average, 20 VDC
+47 dBm (50 W) peak for 1
μ
s
C A U T I O N
The U2004A USB power sensor is DC coupled. DC coupling of the input allows excellent
low frequency coverage. However, the presence of external DC component signals may
affect the accuracy of the power measurement. Users are advised to use suitable external
DC block for the DC component removal.
DC voltages in excess of the maximum value (5 V) can damage the sensing diode.