LED Indicator Guide
The LED indicator is found at the rear panel of the power sensor. The
following table shows the states of the LED indicator and its description.
Table 1-1
States of LED indicator and its description
LED indicator
Description
GREEN blinking
Device in communication mode.
Example: Sending SCPI commands or reading data.
RED blinking
Error - Highest priority event
The error is due to HW/OS/Self-test. All other LED indicators will
not function while having this error. The error message can be read
by sending SYST:ERR? command. Users are advised to return the
power sensor to Agilent.
RED
Error. Use SYST:ERR? to read the error message. Reading the error
message will also clear the LED indication. Users are advised to read
the message as some of the error might cause measurement errors.
Example of error cause:
1
SCPI command syntax error
2
Invalid Zero
AMBER
Zeroing is in progress.
Sending SCPI commands during the zeroing process will cause error.
This will cause the LED indicator to turn RED.
Summary of Contents for U2000 Series
Page 1: ...Agilent Technologies Agilent U2000 Series USB Power Sensors Operating and Service Guide...
Page 14: ...xiv U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 16: ...xvi U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 18: ...xviii U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...