9-4
Agilent U1840A Medalist sj5000 Hardware Manual
9
Downloading PLC Programs
Downloading Using the Serial Cable
The RS232 cable is supplied with the sj5000.
The download
procedure is as follows:
1
Connect the RS232C cable between the PC (COM port 1) and
port 1 on the PLC.
2
On the control panel move PLC dip switch 4 on. (This sets the
PLC serial communications protocol to Auto.)
3
Launch CX-Programmer version 6.1.
4
Click
File
>
Open
to open c:\cpi\data\PLC\sj5000.cxt.
5
In the left pane, right-click
NewPLC1(CP1H) Offline
.
6
In the Change PLC dialog box, set the following:
d
Select
CP1H
for Device Type and click
Settings
. Then in the
Device Type Settings dialog box, select
X
for CPU Type
and click
OK
.
e
Select
Toolbus
for Network Type.
f
Click
OK
.
7
Click
PLC
>
Online
.