Requirements for Injection Port Loading

1.  To ensure proper injection in all wells during your assay, 

each series of ports must contain the same injection 

volume (for example, all Port A injection ports must be 

filled with the same injection volume across the entire XF 

Sensor Cartridge). This applies to Background Correction 

wells also.  

2.  All compounds/reagents should be diluted with the ap-

propriate assay media before being loaded into the sensor 

cartridge, unless noted otherwise. For further details, 

consult the appropriate Agilent Seahorse XF Assay Kit/

Reagent user manual. 

NOTE:

 Serum or BSA containing solutions should not be 

loaded into the ports.

3.  The hydrated XF Sensor Cartridge must remain in the XF 

Utility Plate, placed flat on the work surface throughout the 

loading procedure. Do not lift or angle the XF Utility Plate/

XF Sensor Cartridge away from the work surface while 

loading.

4.  Handle the XFe96 sensor cartridge carefully. Hold the base 

of the XF Utility Plate when transporting an XF Sensor Car-

tridge. To mitigate the accidental leaking of injection solu-

tions prior to starting the assay, load the injection ports in 

close proximity to the Seahorse XFe96/96 Analyzer.

5.  A manual or automatic pipet may be used. Ensure any 

pipet is properly calibrated to accurately dispense volumes 

of 20.0 to 30.0 µL. 

6.  The port loading guide, provided with each XF Sensor 

Cartridge, can ensure the desired port is loaded. Use of 

the port loading guide is encouraged, but not required. The 

top of the XF Sensor Cartridge can be easily marked with 

a laboratory pen to provide a visual guide. Avoid marking 

any barcodes or labels on the XF Utility Plate or XF Sensor 

Cartridge.

Figure 2

 XFe96 Sensor Cartridge orientation.

Seahorse

A

C

E

G

B

D

F

H

Loading the XFe96 Sensor Cartridge with 

injection solutions

NOTE: 

The hydrated XF Sensor Cartridge must remain in the 

XF Utility Plate, flat on the work surface throughout the load-

ing procedure. Do not lift or angle the XF Utility Plate/XF Sen-

sor Cartridge away from the work surface while loading. Hold 

the base of the XF Utility Plate when handling the XF Sensor 

Cartridge to avoid triggering any leaking of solutions from the 

injection ports.

Step 1

Prepare and warm injection solutions to 37°C. 

NOTE:

 Injected solutions should be at pH 7.3 - 7.4 at 37°C 

prior to loading into the injection ports. Diluting compounds 

in Agilent Seahorse XF DMEM, pH 7.4 or XF RPMI, pH 7.4 will 

facilitate accurate pH of injection solutions.

Step 2

Orient the XFe96 Sensor Cartridge as shown in Figure 2. 

Place row labels (lettered A-H) to the left. The triangular notch 

(green arrow) will be in the bottom left-hand corner  

(Figure 2).