Basic Procedure

A key feature of the Agilent Seahorse XFe96/XF96 Analyzer is the ability to inject 

solutions during the assay and observe changes in respiration or glycolytic rates  in 

real time. This is accomplished by pipetting solutions into the Injection Ports on the 

XF Sensor Cartridge prior to beginning an XF assay. This procedure describes the 

injection port loading process which is performed the day of your assay following 

overnight XF Sensor Cartridge hydration.

Recommended injection volume is 20-30 µL for XFe96/XF96 Analyzers.  
Recommended Injection Solution Volumes for 10X dilution upon injection, starting 

with a microplate well volume of 180 µL assay medium (Figure 1):

1.  Port A: 20 µl
2.  Port B: 22 µl
3.  Port C: 25 µl
4.  Port D: 28 µl

The composition and number of ports utilized will depend on your assay design.

Loading the Agilent Seahorse XFe96 

Sensor Cartridge Injection Ports

User Guide

Seahorse

A

C

E

G

B

D

F

H

A

B

C

D

Port designation

for each well

Figure 1. 

Recommended injection volumes and port designation for each well.