95
Chapter 5: Maintenance and troubleshooting
BenchCel Microplate Handling Workstation R-Series User Guide
Hardware problems
About this topic
This topic lists possible hardware problems, the causes of the problems,
and ways to resolve the problems. If you are still experiencing problems
with the BenchCel Workstation after trying the solutions, contact
Velocity11 Technical Support.
Hardware problems
Problem
Cause
Solution
The BenchCel device
does not turn on.
Your lab does not meet the
electrical requirements.
Make sure your lab meets the
electrical requirements. See “Verifying
laboratory requirements” on page 29.
The BenchCel device is not
connected to the power source.
Connect the BenchCel device to the
power source. See “Connecting the
power source” on page 39.
The fuse is blown.
Contact Velocity11 Technical Support.
A hissing sound can be
heard.
A leak is present in the air
connection or inside the device.
Check the air connections at the back
of the device and at the source
(house, cylinder, or pump).
If the connections look fine, the leak
is might be inside the device. Contact
Velocity11 Technical Support.
Oil is present in the
stacker head.
The compressed air is not from an
oil-free compressor and oil has
leaked into the BenchCel device.
Contact Velocity11 Technical Support.
The microplate drops
from the robot
grippers.
The labware definition for the
microplate type might contain
incorrect information.
Check the following in the labware
definition: stacker gripper offset,
robot gripper offset, stacking
thickness, plate thickness, sensor
intensity, and the plate-presence
threshold.
To check or correct these values,
open BenchCel Diagnostics and click
the Labware tab.
Note:
In the Plate offsets (mm) area,
type 8 in the Robot gripper offset and
the Stacker gripper offset boxes. 8 mm
is the recommended offset for most
labware.
If the problem persists, contact
Velocity11 Technical Support.
The microplate drops
or is held loosely by
the stacker grippers.
The stacker gripper pressure is too
low so that the grippers are holding
the microplate too loosely.
Increase the stacker gripper pressure.
See “Adjusting the stacker gripper
pressure” on page 131.
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...