Pump Module User Guide
Maintenance and troubleshooting
28
Pump Module User Guide
Troubleshooting
Locate your problem in the following table and try the solution. If the problem
persists after you try the solutions, contact Automation Solutions Technical
Support.
Inspect the Pump Module
tubing for wear and leaks.
Replace the tubing as
necessary.
Monthly
Tube deterioration, or liquid
fails to pump or fails to drain
properly
IMPORTANT
When the Pump
Module is not in use, lift the
flip- top cover on each pump
head to prevent flattening the
tubing and to help maximize
the tubing life.
Calibrate the Weigh Station, if
applicable.
Bimonthly
Deterioration of liquid- level
accuracy in reservoir
Maintenance task
Schedule
Symptoms
WARNING
Ensure that the Pump Module is turned off before you clean the device.
CAUTION
Use only recommended cleaning materials. Using other cleaning solutions and
materials can damage the device. Do not use abrasive, corrosive cleaning agents. Do not use
metal brushes.
Problem
Potential cause
Solution
Tubing creeps through
the pump head instead
of being secured in
place.
The pump head
clamp is too loose
for the tube size.
Adjust the tubing clamp for
the tube size. See
head tube size adjustment”
on page 29
.
The tubing is crushed
or damaged in the
pump head.
The pump head
clamp is too tight
for the tube size.
Adjust the tubing clamp for
the tube size. See
head tube size adjustment”
on page 29
.