Troubleshooting and Diagnostics
5
General Troubleshooting Guide
•
If none of the above steps occur:
•
5 V is present by observing the rotation of stirrer magnet.
•
12 V is present by observing the rotation of ventilation fans
in the baseplate.
Autosampler Problems
Please contact your local Agilent service representative.
RI Problems
•
Fails to autozero correctly
•
Purge the RI thoroughly.
•
Check the purge column is not blocked (if fitted), by observing a
significant decrease in system pressure when the purge valve opens.
•
Low sensitivity
•
Clean the RI cell thoroughly (for example sonicate in 10 % nitric acid).