50
Installation and Verification Manual
2
Getting Started
Optical In Ports
The Optical In ports accept OTN signals at 2.66 Gb/s and 10.71
Gb/s, SONET OC-1, OC-3, OC-12, OC-48 and OC-192 and SDH
STM-0, STM-1, STM-4, STM-16, STM-64 signals, depending on
the model and options fitted.
52 - 2.66 Gb/s
Selectable optical connector (see table below)
for a 52 Mb/s to 2.66 Gb/s optical input (OC-1, OC-3, OC-12,
OC-48, OTU1/STM-0, STM-1, STM-4, STM-16 signals).
Wavelength 1200 to 1600 nm. Input damage power >+3 dBm;
never exceed maximum input power. The recommended input
power operating level for OC-1, OC-3/STM-0, STM-1 signals is
-33 to -10 dBm and for OC-12/STM-4 and OC-48/STM-16/OTU1
signals is -28 to -8 dBm.
9.95/10.71 Gb/s
Selectable optical connector (see table below)
for 9.95/10.71 Gb/s (OC-192/STM-64/OTU2) optical input
signals. Wavelength 1290 to 1565 nm. Input damage power
>+1 dBm; never exceed maximum input power. The
recommended input power operating level for
OTU2/OC-192/STM-64 signals is -18 to -7 dBm for 9.95 Gb/s and
-16 to -7 dBm for 10.71 Gb/s.
Clock Ports
TX Eye Clock 9.95 Gb/s and 10.71 Gb/s
SMA connector
providing a TX Eye Clock signal (at 1/16 of the line rate) which
can be used to trigger an oscilloscope when examining data
signals.
Connector
Option
FC/PC Adapters fitted on all optical interfaces
609
SC Adapters fitted on all optical interfaces
610
ST Adapters fitted on all optical interfaces
611
panther3_iv.book Page 50 Wednesday, January 15, 2003 12:03 PM