Performance Tests
3
Installation and Verification Manual
115
Equipment Required
1
Recall the OmniBER OTN’s default settings as shown on
2
Press
<Menu>
, select
Tx/Rx > Transmitter Settings
and set
the
Signal Mode
to
PDH/DSn
. Now set the
Signal Rate
to
DS1
as shown in the example below.
Oscilloscope
54810A with Option 100
Communication Mask Kit (E2625A)
Patch Cable
15512A Siemens 3 pin
Patch Cable
15670A Bantam 110 ohm
Adapter
E2621A Termination Adapter Probe
N O T E
This performance test uses the Infinium 54810A Oscilloscope with
Option 100 (E2625A) Telecom Masks Template Kit. If any other
Oscilloscope is used refer to the scope information for Mask
measurement. It is assumed the appropriate Masks are installed in the
Infinium.
To test the DS1 output pulse mask
panther3_iv.book Page 115 Wednesday, January 15, 2003 12:03 PM