Running Samples
94
NovoCyte Quanteon Flow Cytometer Operator's Guide
Figure 62.
Compensate FITC Spillover into PE Channel in Spillover Matrix
Figure 63.
Compensate PE Spillover into FITC, PerCP and APC Channels
NOTE
Select
Preview
in the compensation window to visualize the compensation results
dynamically when adjusting the spillover matrix coefficient. Click the
Expand
Statistics
icon in the bottom-right corner of the plot to show the statistical
calculation results. Look at the corresponding Mean or Median value to judge if the
correct compensation has been achieved.
Summary of Contents for NovoCyte Quanteon