Chapter 3
93
Technical Specifications
Noise Figure and Gain
Instrument's own
Noise Figure
Figure 3-1
Characteristic
1
Noise Figure at 23
°
C ± 3
°
C (10 MHz to 3.0 GHz)
Frequency
Noise Figure
Noise Figure over a limited
temperature range of 23
o
C ± 3
o
C
10 MHz to
3.0 GHz
< 4.8 dB + (0.00124 * freq in MHz)
< 4.4 dB + (0.00117 * freq in MHz)
> 3.0 GHz to
13.2 GHz
< 12.0 dB
< 10.5 dB
>13.2 GHz to
26.5 GHz
< 16.0 dB
< 12.5 dB
1. Characteristic values are met or improved on by 90% of instruments with 90% confidence.