HCOPy Subsystem
The HCOPy commands are used to control the printing function.
HCOPy Subsystem Command Reference
Print Screen Image
HCOPy:SDUMp
This command executes to print a screen image.
SYSTem Subsystem
The SYSTem commands are used to control the overall system
function.
SYSTem Subsystem Command Reference
System Error
SYSTem:ERRor?
This query command returns a set of values
<int 1>
and
<string 2>
, for error numbers and error messages. Refer to
for the listings of the error messages.
Lock System Panel
SYSTem:KLOCk ON|OFF|1|0
SYSTem:KLOCk?
This command locks the control panel functions for protection.
At preset, this is set to
OFF
(0).
If this is set to
ON
(1), all of the function become inactive, but
only the
Local
softkey is defined. When the
Local
softkey is
pressed or the command
SYSTem:KLOCk OFF
is sent, the
Tester returns to the normal operation.
This command is not identical to the panel key lock on the
[Configuration] screen. See