Programming Command Reference
5
N9360A Multi UE Tester W-CDMA Programming Manual
5-61
•
<real 3>
(up to 6 letters) is the measured data ranging from
-99.99
to
0.00
in dB. If measurement result is out of range
or Origin Offset is not measured, the value
-99.99
is
returned.
BER Measurement
TESTs:MANual:RESults:BER?
This query command returns the BER measurement result values,
<int 1>
,
<int 2>
, and
<real 3>
. Each value is separated by a
comma.
•
<int 1>
(1 digit) is
0
if measurement result is within the
measurement range, or
1
if measurement result is out of
range or BER is not measured.
•
<int 2>
(1 digit) is the test result for the pass (0) or fail (1).
•
<real 3>
(up to 7 letters) is the measured data ranging from
0.00
to
-100.00
in %. If measurement result is out of range
or BER is not measured, the value
+99.99
is returned.
CPICH RSCP Measurement
TESTs:MANual:RESults:CPichRScp?
This query command returns the CPICH RSCP measurement
result values,
<int 1>
,
<int 2>
, and
<int 3>
. Each value is
separated by a comma.
•
<int 1>
(1 digit) is
0
after measurement, or
1
if CPICH RSCP
is not measured.
•
<int 2>
(1 digit) is the same value as that of
<int 1>
.
(Decision with limits is not made.)
•
<real 3>
(up to 6 letters) is the measured data, or
-99
if
CPICH RSCP is not measured.
Command Reference for MS Information
The following MS Information commands are used to get
information on the mobile phone under test.
Summary of Contents for N9360A
Page 1: ...Agilent Technologies Agilent N9360A Multi UE Tester W CDMA Programming Manual...
Page 4: ...1 4 N9360A Multi UE Tester W CDMA Programming Manual 1 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 8: ...8 N9360A Multi UE Tester W CDMA Programming Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 12: ...6 N9360A Multi UE Tester W CDMA Programming Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...