Functions and Measurements
3
N9342C/43C/44C User’s Guide
75
Channel Scanner Setup
To customized the channel scan process, you can
press
[Meas] > {Channel Scanner} > {Setup}
to set the
measurement interval and rule in the submenu.
The
Meas Interval
is a value of time/distance during
which the every two scan process is triggered.
When it is set to off, the channel will be scanned
with default settings.
The
Interval Type
is toggled between time and
distance. When it is set to
Time
, the minimum
interval time is 100 ms. When it is set to
Distance
,
it allows you to measure the channel with the
specific distance on the road test. The minimum
interval distance is 1 meter. Please refer to the
on page 80 for the distance interval
channel scan.
The
Measure Rule
and
Search Rule
is toggled between
Speed
and
Accuracy
.
For each scan process, the analyzer will first
search and find out the available channels
according to the selected scan mode. Then, a
precise measurement will perform to get the scan
result.
The accuracy mode in measurement rule is set as
default to get the precise measurement for the
searching- out channels.
The speed mode will speed the measurement but
reduce the measurement accuracy. It is set to
default in search rule to search out the available
channels approximately.
The
Top Cycle
number is the searching times for
each measurement. It allows you to measure the
specific searching- out channels continuously.
NOTE
The option GPS is required to set the Interval Type to
Distance. Press [System] > {More} > {GPS} > {GPS On}
Summary of Contents for N9342C
Page 1: ... Agilent N9342C 43C 44C Handheld Spectrum Analyzer User s Guide ...
Page 8: ......
Page 39: ...31 AgilentN9342C 43C 44C Handheld Spectrum Analyzer 3 Functions and Measurements ...
Page 92: ...3 Functions and Measurements 84 N9342C 43C 44C User s Guide ...
Page 156: ...5 Error Messages 148 N9342C 43C 44C User s Guide ...