Chapter 6
Spurious Emissions Measurement
Measurement Procedure
Measurement Procedure
Step 1.
Press
Mode
,
TD-SCDMA with HSPA/8PSK
to enable the TD-SCDMA with
HSPA/8PSK mode measurements.
Step 2.
Press
Mode Preset
to preset the analyzer.
Step 3.
Press
Input/Output
,
External Gain
,
BTS
and enter the number and
−
dB
according to the external attenuator connected to the analyzer.
Step 4.
Press
Input/Output
,
RF Input
,
RF Coupling
to toggle the RF Coupling to
DC
.
Step 5.
Press
Mode Setup
,
Radio
,
Device
to toggle the device to
BTS
.
Step 6.
Press
Meas
,
Spurious Emission
to initiate the spurious emission
measurement.
Step 7.
Press
Meas Setup
,
More
,
Range Preset
,
Freq Setup
and enter the
appropriate carrier parameters for your testing
Step 8.
Press
Return
to back to Range Preset menu and select
Category A
or
Category B
for your testing. Range Table parameters and limit lines will
be set automatically according to your selection.
Category A or Category B define the Base Station mandatory spurious
emissions limits with the specified frequency ranges. For more
information, see
N9079A TD-SCDMA with HSPA/8PSK User’s and
Programmer’s Reference
.
The
Mobile
key is grayed out when you are in BTS test mode. It is
available only when you select
MS
by pressing
Mode Setup
,
Radio
,
Device
.
Step 9.
Check the spurs reported in the text window and press
Meas Setup
,
Spur
and enter the spur number to highlight the data you are interested in.
The marker in the graph window will move synchronously.
Toggle the
Meas Type
key between
Examine
and
Full
. The
Examine
type
sets the average mode to
Exp
and the
Full
type sets the average mode to
Repeat
.
Step 10.
Press
More
,
Spur Report
and toggle the mode from
All
to
LimitTest
. Only
spurs above the limit line are displayed.
Step 11.
If you want to change the measurement parameters from their default
condition for a customized measurement, press
Meas Setup
,
Range Table
to adjust the parameters for your testing.
For more information about changing measurement parameters, see
N9079A TD-SCDMA with HSPA/8PSK User’s and Programmer’s
Reference
.
Summary of Contents for N9010A
Page 4: ......
Page 7: ...Contents 7 Waveform Signal Envelope Views of I only or Q only 106...
Page 8: ...8 Contents...
Page 12: ...12 Chapter1 Making TD SCDMA with HSPA 8PSK Measurements Setting Up and Making a Measurement...
Page 25: ...Chapter 3 25 ACP Measurements Measurement Procedure Messages Guide...
Page 28: ...28 Chapter3 ACP Measurements Measurement Results...
Page 44: ...44 Chapter5 Spectrum Emission Mask Measurements Troubleshooting Hints...
Page 64: ...64 Chapter8 Power Statistics CCDF Measurements Troubleshooting Hints...
Page 94: ...94 Chapter12 Monitor Spectrum Measurements Measurement Results...