40
Chapter 5
Spectrum Emission Mask Measurements
Measurement Procedure
Measurement Procedure
Step 1.
Press
Mode
,
TD-SCDMA with HSPA/8PSK
to enable the TD-SCDMA with
HSPA/8PSK mode measurements.
Step 2.
Press
Mode Preset
to preset the analyzer.
Step 3.
Press
Input/Output
,
External Gain
,
BTS
and enter the number and
−
dB
according to the external attenuator connected to the analyzer.
Step 4.
Press
Mode Setup
,
Radio
,
Device
to toggle the device to
BTS
.
Step 5.
Press
Mode Setup
,
Analysis Timeslot
to select the timeslot to be
measured.
Step 6.
Press
Sweep/Control
,
Gate
and toggle the
Gate
to
On
. Press
More
,
Gate
Source
and select the desired gate source
External 1
.
If no measurement is performed and only a symbol of yellow ‘*’ displays
on the top right of the screen, you need to check the gate settings. You
need the correct gate signal to make the SEM measurement.
Step 7.
Press
FREQ Channel
,
2
,
GHz
to set the center frequency to 2 GHz.
Step 8.
Press
Meas
,
Spectrum Emission Mask
to initiate the Spectrum Emission
Mask measurement.
Step 9.
Press
Meas Setup
to check
the keys available to change the measurement
parameters from the default condition.
To make an averaged measurement, press
Avg/Hold Num
to toggle the
state to
On
and enter the number of averages.
To select the power reference, press
Meas Type
key to select from
Total
Power Ref, PSD Ref
and
Spectrum Pk Ref
.
To adjust the offset parameters, toggle the
Limit State
key to
Manual
and
press
Offset/Limits
to change the offset parameters and limit lines.
Step 10.
Press
View/Display
and select the
desired view from
Abs Pwr Freq
,
Rel Pwr
Freq
and
Integrated Power
. You can also turn on or turn off the limit lines by
toggle the
LimitLines
between
On
and
Off
.
For more details about changing measurement parameters, see
N9079A TD-SCDMA with HSPA/8PSK User’s and Programmer’s
Reference
.
If you have a problem, and get an error message, see the “
Instrument
Messages Guide
”.
Summary of Contents for N9010A
Page 4: ......
Page 7: ...Contents 7 Waveform Signal Envelope Views of I only or Q only 106...
Page 8: ...8 Contents...
Page 12: ...12 Chapter1 Making TD SCDMA with HSPA 8PSK Measurements Setting Up and Making a Measurement...
Page 25: ...Chapter 3 25 ACP Measurements Measurement Procedure Messages Guide...
Page 28: ...28 Chapter3 ACP Measurements Measurement Results...
Page 44: ...44 Chapter5 Spectrum Emission Mask Measurements Troubleshooting Hints...
Page 64: ...64 Chapter8 Power Statistics CCDF Measurements Troubleshooting Hints...
Page 94: ...94 Chapter12 Monitor Spectrum Measurements Measurement Results...