4-20
Service Guide N5242-90001
Troubleshooting
PNA Series Microwave Network Analyzers
Rear Panel Troubleshooting
N5242A
C:>ping 141.121.69.162
Pinging 141.121.69.162 with 32 bytes of data:
Reply from 141.121.69.162: bytes=32 time<10ms TTL=127
Reply from 141.121.69.162: bytes=32 time<10ms TTL=127
Reply from 141.121.69.162: bytes=32 time<10ms TTL=127
Reply from 141.121.69.162: bytes=32 time<10ms TTL=127
Ping statistics for 141.121.69.162:
Packets: Sent = 4, Received = 4, lost = 0 <0% loss>.
Approximate round trip times in milli-seconds:
Minimum = 0ms, Maximum = 0ms, Average = 0ms
6. The above message verifies that one way communication from the analyzer to the
network has been established
7. If the subnet mask was changed in step 2, set it back at this time.
How to Ping from the Local Area Network (LAN) to the Analyzer
Reverse communication should also be verified. Determining this, though, is dependent
upon your network setup and software. Generally, you need to issue a ping command using
the IP address of the analyzer to be tested. For example, using Windows 95, 98, 2000, or
XP and while at a DOS prompt, type in
ping xxx.xxx.xxx.xxx
1
. Then press ENTRY
on the front panel. If full communication can be established, then the computer
display shows the cycle time for each of four cycle attempts (similar to that in step 5).
Other software may behave somewhat differently, but basically the same.
If the analyzer can talk to the network, but the network can not talk to the analyzer, then the
computer or device used from the network may have a subnet mask that excludes
communication with the IP address chosen for the analyzer. Any subnet mask other than
0.0.0.0
will exclude operation from some addresses. Changing the subnet mask of a computer
or other device should only be attempted by a qualified network administrator. Failure to
communicate due to a subnet mask incompatibility does not indicate any failure of the analyzer.
If the analyzer fails to ping in either direction, and assuming the subnet masks are set
properly, then the fault must be isolated to the analyzer or to the network. Contact a
qualified network administrator.
Testing Between Two Analyzers
The ability of the analyzer's LAN to function can be easily tested by connecting two
analyzers together using a “crossover cable” (a short length of cable with an RJ-45
connector on each end).
Some network hubs have the capability to make a crossover connection using two normal,
or straight-through, cables. If this capability is not available and a crossover cable is not
available, a crossover cable can be made by following the directions in
1.
The letters x represent the IP address of the analyzer.
Enter
Summary of Contents for N5242A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 33: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 75: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 119: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 153: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 237: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 303: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...