7-8
Service Guide N5242-90001
Repair and Replacement Procedures
PNA Series Microwave Network Analyzers
Removing and Replacing the Front Panel Assembly
N5242A
Removing and Replacing the Front Panel Assembly
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16 inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
for this procedure.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
“Removing the Covers” on page 7-6
3. With a 5/16 inch torque wrench, remove all the semirigid jumpers (item
①
) from the
front panel.
4. With a T-10 TORX driver, remove the 12 screws (item
②
) from the sides of the frame.
CAUTION
Before removing the front panel from the analyzer, lift and support the front
of the analyzer frame.
5. Slide the front panel over the test port connectors.
6. Disconnect the ribbon cable (item
③
) from the A1 front panel interface board.
Replacement Procedure
IMPORTANT
When reconnecting the front-panel jumpers, torque the connectors to 10 in-lb.
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Summary of Contents for N5242A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 33: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 75: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 119: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 153: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 237: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 303: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...