3-40
Service Guide N5242-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
Adjustments
N5241A/42A
Receiver Adjustment
The receiver calibration is used to adjust the network analyzer receivers for a flat response
across its full frequency range:
1. A power meter/sensor is connected to Port 1, as shown in
, to establish a
reference for flatness.
2. A cable is inserted between the power sensor and the test port, as shown in
to establish a reference for the cable.
3. The same cable is connected between test port 1 and test port 2, as shown in
, and a signal from Port 1 is used to adjust the “B” receiver at Port 2.
The adjustment is repeated using a signal from Port 2 to adjust the “A” receiver at
Port 1.
Data obtained during this adjustment are stored in the
mxcalfile_pxx
files on the hard disk
drive. The data are used in subsequent measurements.
If the hard disk drive is replaced, these
mxcalfile_pxx
files will be lost. Therefore, they
should be backed up (saved on a floppy disk) so that they can be restored. If using multiple
disk drives (e.g. classified and general usage), then these files must be replaced on each
individual disk drive.
These files can be recreated by performing another receiver calibration adjustment.
Equipment Used for the Receiver Adjustment
Procedure
1. Connect the equipment as shown in
. Connect a GPIB cable between the
network analyzer and the power meter.
Equipment Type
Model or
Part Number
Alternate Model
Part Number
Power meter
E4418B/E4419B
E4418A/E4419A
Power sensor, 3.5 mm
E4413A
8485A
Adapter, 3.5 mm (f) to 3.5 mm (f)
83059B
85052-60012
RF Cable, 3.5 mm (f) to 3.5 mm (f)
85131C
85131E
Summary of Contents for N5241A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 35: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 81: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 125: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 159: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 243: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 313: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...